ICP ANALYSIS OF METAL/METALLOID PARTICULATES FROM SOLDER OPERATIONS
Method Number: |
ID-206 |
|
Matrix: |
Air, Wipe (Smear Tab), or Bulk |
|
OSHA Time Weighted Average (TWA)
Permissible Exposure Limits: |
0.01 mg/m3 Silver (Ag)
0.002 mg/m3 Beryllium (Be)*
0.1 mg/m3 Cadmium (Cd) fume
0.1 mg/m3 Copper (Cu) fume
0.05 mg/m3 Lead (Pb)
0.5 mg/m3 Antimony (Sb)
2.0 mg/m3 Tin (Sn)
5.0 mg/m3 Zinc Oxide (ZnO) fume |
|
Collection Procedure: |
A personal sampling pump is used to draw a known
volume of air through a mixed cellulose ester
membrane filter contained in a polystyrene cassette.
Wipe (smear tab) and bulk material are collected by
grab sampling techniques. |
|
Recommended Sampling Rate: |
2 liters per minute |
|
Recommended Air Volumes
Be, Cd, Cu, Pb, Sb, Sn, ZnO:
If Ag is determined: |
480 L 960 L |
|
Analytical Procedure: |
Filters are digested with hydrochloric and nitric acids.
Analysis is performed using Inductively Coupled Plasma-Atomic Emission Spectroscopy. |
|
Detection Limits: |
See Section 2 |
|
Validation Range: |
See Section 3 |
|
Method Classification: |
Validated analytical method |
|
Chemist: |
Dixon C. Cook |
|
Date: |
May, 1991 |
* |
This method may be used to determine STEL (0.005 mg/m3) or Ceiling (0.025 mg/m3)
exposures to Be. At least a 30-min sample (2 L/min) should be taken for STEL determinations; at least 5-min for Ceiling monitoring. |
Commercial manufacturers and products mentioned in this method are for
descriptive use only and do not constitute endorsements by USDOL-OSHA.
Similar products from other sources can be substituted.
Division of Physical Measurements and Inorganic Analyses
OSHA Technical Center
Salt Lake City, Utah
1. Introduction
1.1. Scope
1.1.1. This method describes the collection and analysis of airborne metal and metalloid
particulates from solder operations in industry. Time Weighted Average (TWA) air
samples are collected using personal sampling pumps and mixed-cellulose ester (MCE)
filters. Analysis is by Inductively Coupled Plasma-Atomic Emission Spectroscopy
(ICP-AES). Bulk and wipe samples can also be collected and analyzed by this method.
1.1.2. This method provides rapid simultaneous analysis and data reduction for a wide range of
elements, eliminating the necessity of separate analyses by conventional atomic absorption
techniques.
1.1.3. This method was validated for 8 elements (Ag, Be, Cd, Cu, Pb, Sb, Sn, and Zn). A total
of 21 elements are analyzed, but 13 of these are determined for screening purposes only
(Al, As, Ca, Co, Cr, Fe, Ni, Mg, Mn, Mo, Se, Si, and V). Other elements can be added
to or subtracted from the method. The capability for expanding the analysis to other
elements is dependent on laboratory instrumentation, and element solubility and stability
in the acid matrix used for digestion.
1.1.4. The elements validated are those commonly found in solders or soldering operations. The
acid matrices used for sample digestion (4:1 ratio of HCl:HNO3) and dilution (32%
HCl/4% HNO3) were selected to ensure solubility of the eight elements.
1.2. History
1.2.1. Previous to the introduction of ICP-AES, samples containing metallic particulates were
digested in a variety of ways and analyzed by Atomic Absorption Spectroscopy (AAS) at
the OSHA Salt Lake City Analytical Laboratory (SLCAL).
1.2.2. An ICP simultaneous spectrometer [Instruments SA (ISA), Model JY-32, Edison, NJ] was
used to validate this method.
This method is applicable to any simultaneous spectrometer. The validation of the JY-32
is described in the backup report (8.1.).
2. Detection Limits and Working Ranges (8.1.)
2.1. Detection limits are listed below:
|
|
Qualitative Detection Limit |
Quantitative Detection Limit |
|
Element |
(µg/mL) |
(µg*) |
(µg/mL) |
(µg*) |
|
Ag |
0.018 |
0.45 |
0.061 |
1.5 |
Be |
0.00029 |
0.0072 |
0.00086 |
0.022 |
Cd |
0.0062 |
0.15 |
0.0205 |
0.51 |
Cu |
0.0078 |
0.19 |
0.026 |
0.65 |
Pb |
0.071 |
1.8 |
0.237 |
5.9 |
Sb |
0.14 |
3.5 |
0.47 |
12. |
Sn |
0.074 |
1.8 |
0.246 |
6.1 |
Zn |
0.0075 |
0.19 |
0.025 |
0.63 |
|
* 25-mL solution volume |
|
The quantitative limit is considered as the lower working range limit. These values are based on
instrument performance and will normally change over time.
2.2. The upper working range limit is 100 µg/mL for all validated elements except Be. Due to the high
sensitivity resulting from instrument settings and choice of wavelength, the upper limit for Be is 10 µg/mL.
3. Precision and Accuracy (8.1.)
3.1. The precision and accuracy data for the 8 validated elements are listed below:
|
Element |
Range
(µg) |
Bias |
CV1 (pooled) |
Overall Error (±%) (Total) |
|
Ag |
4.6 |
- |
19. |
0.006 |
0.054 |
11.5 |
Be |
0.47 |
- |
2.0 |
-0.075 |
0.037 |
14.9 |
Cd |
23. |
- |
100. |
0.048 |
0.041 |
13.1 |
Cu |
23. |
- |
94. |
0.065 |
0.038 |
14.2 |
Pb |
11. |
- |
48. |
-0.038 |
0.060 |
15.7 |
Sb |
126. |
- |
510. |
0.014 |
0.043 |
10.1 |
Sn |
476. |
- |
2,000. |
-0.008 |
0.043 |
9.4 |
Zn |
973. |
- |
2,900. |
0.031 |
0.039 |
10.9 |
|
These values are based on six samples at each of three concentration levels tested. Solutions of the
eight elements were spiked on MCE filters. These samples were then digested and analyzed as in
Section 6. of this method.
3.2. The eight elements reported above were spiked at 0.5, 1, and 2 times the PEL, assuming a 480-L
air volume for all elements except Ag where a 960-L air volume was used. Overall analytical
errors for the elements tested were within ±25%; the greatest value was 15.7% for Pb. The large
overall error may be attributed to the fact that Pb was validated near its detection limit.
4. Interferences (8.2.)
High temperatures present in the plasma (5,000 to 8,000 °C) minimize most chemical and matrix
interferences. Interferences do exist, however, and can be categorized as follows:
4.1. Physical interferences such as nebulization and transport effects are influences that determine the
rate and particle size in which analytes are delivered to the plasma. These effects are minimized
by matching the acid concentrations of samples and standards.
4.2. Chemical interferences are characterized by molecular compound formation, ionization effects, and
solute volatilization effects. These effects are not severe in ICP analysis and are minimized by
matrix matching and careful selection of operating conditions such as: Incident plasma source
power, sample uptake rate, and plasma observation height.
4.3. Spectral interferences include:
- Unresolved overlap of molecular band spectra.
- Overlap of a spectral line from another element.
- Background from continuous or recombination phenomena.
- Background from stray light.
The first effect (a) can be minimized by a careful selection of wavelengths for the reported
elements. The wavelengths selected for analysis using the JY-32 are shown in Table 1. The other
types of spectral interferences (spectral overlap and elevated background) are minimized by software
which performs interelement corrections. This software assumes a linear relationship between the
analyte concentration and interference within the working
range limits. The spectral interference correction equation typically used by ICP manufacturers is:
Corrected Concn = Calculated Concn - Ai × CPi
Where:
Ai = Correction factor
CPi = Concentration of the interfering element
Samples having analyte concentrations above the working range limits listed in Section 2.2. should
be diluted into range; interelement corrections may not be accurate above the working range.
Further information and experimentally determined interelement corrections for the validated
elements are listed in the backup report (8.1.).
4.4. If necessary, supplemental background correction can be performed with additional software
supplied by the instrument manufacturer.
5. Sampling
5.1. Safety Precautions
5.1.1. Attach the sampling equipment to the worker such that it will not interfere with work
performance or safety.
5.1.2. Follow all safety practices that apply to the work area being sampled.
5.1.3. Wear impermeable gloves when taking wipe or bulk samples.
5.2. Equipment
5.2.1. Mixed cellulose ester (MCE) filters (0.8-µm pore size), cellulose backup pads, and twoor
three-piece cassettes, 37-mm diameter, (part no. MAWP 037 A0, Millipore Corp.,
Bedford, MA).
5.2.2. Gel bands (Omega Specialty Instrument Co., Chelmsford, MA) for sealing cassettes.
5.2.3. Sampling pumps capable of sampling at 2 liters per minute (L/min).
5.2.4. Assorted flexible tubing.
5.2.5. Stopwatch and bubble tube or meter for pump calibration.
5.2.6. Scintillation vials, 20 mL, (part no. 74515 or 58515, Kimble, Div. of Owens-Illinois Inc.,
Toledo, OH) with polypropylene or Teflon cap liners. If possible, submit bulk or wipe
samples in these vials for ICP analysis.
5.2.7. Smear tabs, (part no. 225-24, SKC Inc., Eighty Four, PA) for wipe sampling.
5.2.8. Gloves, disposable (for wipe sampling).
5.3. Sampling Procedure - Air Samples
5.3.1. Place a MCE filter and a cellulose backup pad in each twoor three-piece cassette.
Seal each cassette with a gel band.
5.3.2. Calibrate each personal sampling pump with a prepared cassette in-line to within ±10%
of the recommended flow rate of 2 L/min.
5.3.3. Attach prepared cassettes to calibrated sampling pumps (the backup pad should face the
pump) and place in appropriate positions on the employees or the workplace areas.
Whenever possible, collect samples using the air volumes listed below:
Suspected Analyte |
Recommended Air Volume |
|
All analytes except Ag |
480 L |
If Ag is also suspected |
960 L |
STEL or Ceiling for Be |
60 L (STEL), 10 L (Ceiling) |
5.3.4. If the filter becomes overloaded with particulate while sampling, another filter cassette
should be prepared. Consecutive samples using shorter sampling periods should be taken
if overloading occurs.
5.3.5. Place plastic end caps on each cassette after sampling.
5.3.6. Attach an OSHA-21 seal around each cassette in such a way as to secure the end caps.
5.4. Sampling Procedure - Wipe Samples
5.4.1. Wear clean, impervious, disposable gloves when taking each wipe sample to prevent
sample contamination.
5.4.2. Moisten the wipe filters with deionized water prior to use.
5.4.3. If possible, wipe a surface area covering 100 cm2.
5.4.4. Fold the wipe sample with the exposed side in.
5.4.5. Transfer the wipe sample into a 20-mL scintillation vial and seal with vinyl or electrical
tape. Securely wrap an OSHA-21 seal length-wise from vial top to bottom.
5.5. Sampling Procedure - Bulk Samples
In order of laboratory preference, bulk samples may be one of the following:
- a high-volume filter sample,
- a representative settled dust (i.e. rafter) sample,
- a sample of the bulk material in the workplace.
Transfer the bulk material into a 20-mL scintillation vial and seal with vinyl or electrical tape.
Securely wrap an OSHA-21 seal length-wise from vial top to bottom.
5.6. Shipment
5.6.1. When other compounds or elements are known or suspected to be present in the sampled
air, such information should be transmitted with the sample.
5.6.2. Request ICP/Solder analysis. If silver is suspected in any of the samples, request
ICP/Solder-Silver analysis.
5.6.3. Submit at least one blank sample with each set of air or wipe samples. Blank filter samples
should be handled in the same manner as other samples, except an actual sample is not taken.
5.6.4. Send the samples to the laboratory with the OSHA-91A paperwork.
5.6.5. Ship bulk samples separately from air samples. They should be accompanied by Material
Safety Data Sheets, if available. Check current shipping restrictions and ship to the
laboratory by the appropriate method. The type of bulk sample should be stated on the
OSHA-91A and cross-referenced to the appropriate air sample(s).
6. Analysis
6.1. Safety Precautions
6.1.1. Handle reagents and samples carefully. Use protective equipment such as: gloves,
laboratory coats, safety glasses, and an exhaust hood. Wear a fit-tested
respirator if necessary. Clean up spills immediately.
6.1.2. Carefully prepare any solutions or samples (especially those containing acid) within a
suitable ventilated exhaust hood. The acids used to digest the samples are very corrosive
and volatile.
6.1.3. Do not view the plasma directly.
6.1.4. Do not override the radio-frequency generator or torch box safety interlocks.
6.2. Equipment
6.2.1. Inductively coupled plasma/atomic emission direct-reading spectrometer, nebulizer, cooling
unit for torch assembly, computer, and radio-frequency (rf) generator.
6.2.2. Automatic sampler.
6.2.3. Peristaltic pumps (optional). Use one pump for automatic sampler rinse. Use the other
pump for sample introduction into the nebulizer.
6.2.4. Mass flow controller (optional). Use the controller to regulate nebulizer argon flow and
sample uptake rate.
6.2.5. Borosilicate glass Phillips beakers, 125 and 250-mL.
6.2.6. Borosilicate glass volumetric flasks, 10-, 25-, 100-mL, and 1-L or 2-L. Use the larger
flasks to prepare standard solutions.
6.2.7. Hot plate capable of reaching 300 °C.
6.2.8. Mixed cellulose ester filters (0.45-µm pore size) and a filtering apparatus.
6.2.9. Amber colored bottle (1-L) with screw-top cap and Teflon cap liner.
6.2.10. Volumetric pipets, glass. Various sizes for sample dilutions and standard preparation.
6.2.11. Analytical balance (0.01 mg).
6.3. Reagents (Reagent grade or better)
Please see precautions in Sections 6.1.1.-6.1.2. before preparing any reagents having an acid
matrix.
6.3.1. Deionized water (DI H2O).
6.3.2. Hydrochloric acid (HCl), concentrated (36.5 to 38% w/w).
6.3.3. Nitric acid (HNO3), concentrated (69 to 71% w/w).
6.3.4. Reagent blank - sample diluting solution (32% HCl/4% HNO3 by volume).
Prepare as described below:
- Slowly and carefully add 320 mL concentrated HCl to 500 mL DI H2O in a 1-L
volumetric flask. Gently swirl the solution and let cool.
- Slowly and very carefully add 40 mL concentrated HNO3 to the flask,
gently swirl, and let cool.
- Dilute to volume with DI H2O. Mix thoroughly.
6.3.5. Stock solutions of 1,000 µg/mL for the various elements.
6.3.6. Argon - grade or quality specified by the ICP instrument manufacturer.
6.4. Standard Preparation
Prepare multielement working and control standard solutions using 1,000 or 10,000 µg/mL stock
solutions. Multielement standard concentrations recommended for the JY-32 are shown in
Table 1. A suggested control standard is shown in Table 2. Whenever possible, prepare the control
standard from different stock solutions than those used for calibration standards. The control
standard should contain elements and concentrations reflecting what is expected in the majority of
the samples, or problem elements. Any standards containing silver should be prepared in amber
colored containers to protect the silver from exposure to light.
The final acid concentration of the working and control standards is 32% HCl/4% HNO3.
These standards should be stable for at least a year. Also prepare a reagent blank (32% HCl/4% HNO3).
6.5. Sample Preparation
The final acid concentration for the different sample matrices should be 32% HCl/4% HNO3.
All of the elements validated are soluble when using the following acid digestion procedures. Filters,
backup pads, wipes, and bulks are prepared by the following procedures:
6.5.1. Mixed cellulose ester membrane filters, wipe samples (smear tabs), and contaminated
back-up pads.
- Acid-wash the insides of the 125-mL Phillips beakers by refluxing 1:1
HNO3 using a hot plate in a ventilated hood.
Remove the beakers from the hotplate and carefully pour the used 1:1
HNO3 into an appropriate labeled container.
Allow beakers to cool, then rinse several times with DI H2O
and allow to dry.
- Place sample filters in separate labeled and washed beakers.
Note: |
Do not combine sample filters (i.e. if an industrial hygienist has taken two consecutive
samples from an employee). Always prepare each filter separately. If necessary,
combine the results of consecutive samples. |
- If the backup pad has been contaminated during sample collection, digest the pad
in a separate beaker. Also separately digest and analyze a blank backup pad.
- Slowly add 8 mL HCl to each beaker and allow to sit for 5 minutes while swirling
gently. Add 2 mL HNO3.
Note: |
Always add the HCl first to prevent conversion of Sn to
SnO2 or Sb to Sb2O3.
Always slowly add the HCl and allow the filter to become saturated before adding the
HNO3. Loss of Sb or Sn is possible if the procedure is not
strictly followed. See reference 8.1. for further details. |
- Digest the filters in the beakers on a hot plate until the solution volume is approximately:
Solution volume |
Air volume |
|
1 mL |
> 400 L |
0.5 mL |
< 400 L or if Ag is requested |
Note: |
Do not allow the solutions to evaporate to dryness. Loss of analyte may occur. |
Remove the beakers from the hotplate and allow to cool in the exhaust hood.
- For samples with air volumes > 400 L:
Carefully add 8 mL HCl to the beaker, and gently swirl. Quantitatively transfer
the sample to a clean 25-mL volumetric flask with DI H2O rinses.
Be sure to rinse the sides of the beaker. Dilute to volume with DI H2O.
- For samples with air volumes < 400 L or if Ag is a requested analyte:
Add 3 mL of HCl to the beaker and gently swirl. Quantitatively transfer the
sample with DI H2O rinses to a 10-mL volumetric
flask and then dilute to volume with DI H2O.
- If particulate matter is present after dilution, filter the solution through a 0.45-µm
MCE filter contained in a filtering apparatus. Save the filtrate. Repeat the
digestion procedure above for the filter containing the particulate.
6.5.2. Polyvinyl Chloride (PVC) Filter Samples
Note: |
Polyvinyl chloride filters are not routinely used for solder sample collection and analysis
by ICP. In some cases an industrial hygienist will sample for total or respirable dust using
PVC filters and also submit these samples for ICP analysis. The PVC filter will not be
completely digested using this procedure. |
For PVC filters, proceed as in Section 6.5.1., Step 1. Since the PVC filter will not
completely digest, thoroughly rinse the filter residue with DI H2O during quantitative
transfer of the sample solution.
6.5.3. Bulk samples
- Weigh a representative portion of the bulk sample and transfer to an acid-washed
250-mL Phillips beaker.
Note: |
Aliquot amounts of bulks are dependent on the analytical sensitivity,
detection limit, and solubility of the material used. If uncertain, a 20 to 50-mg
aliquot of a solid material can be taken as a starting point. Make sure the aliquot
taken is representative of the entire bulk sample. If necessary, use a mortar and
pestle to grind any nonhomogenous particulate bulk samples in an exhaust hood. |
- Slowly add 16 mL HCl and allow to sit for 5 minutes while swirling gently. Then
add 4 mL HNO3.
- Digest the bulk until approximately 4 mL of sample solution remains. Remove
the sample from the hot plate and allow to cool.
- Slowly and carefully add 32 mL of HCl.
- Quantitatively transfer the sample to a 100-mL volumetric flask with
DI H2O rinses, and gently swirl.
Dilute to volume with DI H2O.
- If particulate matter is present after dilution, filter the solution through
a 0.45-µm MCE filter contained in a filtering apparatus.
Save the filtrate. Repeat the digestion procedure above for the filter containing the particulate.
6.6. Instrument Startup and Calibration
Follow the instrument manufacturer's instructions for instrument start-up and calibration. An
example of ICP operating parameters for the JY-32 with a Meinhard-type nebulizer is shown below.
These settings may vary from instrument to instrument:
Gas used |
Argon |
|
Gas flow |
Plasma Nebulizer Auxiliary plasma |
12 to 16 L/min * 0.14 to 0.18 L/min |
|
Power (rf) |
Incident Reflected |
1.1 kW < 5 W |
|
Observation height |
Plasma |
15 mm above work coil |
|
Wash time |
With automatic sampler Without automatic sampler |
60 s 10 s |
|
Number of exposures |
Standards & samples |
5 |
|
Integration time |
Peak signal |
10 s |
|
Nebulizer |
Solution uptake rate Pressure |
0.8 to 1.6 mL/min ~ 30 psig |
|
Mass flow controller |
Flow rate range |
varies* |
|
* |
Nebulizer is controlled by a mass flow controller.
The nebulizer flow will vary depending on the type in use. |
6.6.1. Profile the instrument before beginning the calibration and analysis. Follow the standard
operating procedure (SOP) (8.3.) or manufacturer's instructions for computer initialization
and profiling.
6.6.2. A two-point calibration curve is normally obtained by nebulizing the working standards
into the plasma and measuring atomic emission intensities. For calibrations using the
JY-32, a first-order linear fit of the data is computer calculated
and slope and intercept coefficients are obtained. Calibrations should be performed by following
the instrument manufacturer's guidelines.
6.7. Analytical Procedure
For more details regarding analytical procedures, refer to the SOP (8.3.) or instrument
manufacturer's software manual(s).
Note: |
The samples and standards are made up in 32% HCl/4% HNO3.
These solutions are very caustic. Care should be exercised during analysis to prevent physical contact
with these solutions. |
6.7.1. If necessary, calculate detection limits. The manufacturer's software (if available) should
be used for determining detection limits. These limits normally do not significantly change
during short time spans. A general rule would be to calculate detection limits every 2 to
4 months or when an integral component (nebulizer, torch, mass flow controller, etc.) of
the ICP has been replaced or adjusted. A typical calculation of detection limit (DL) in µg
is shown:
DL = |
(K × SDI × C) (I - Io) |
× S |
Where: |
K |
= |
Degree of confidence (sigma value)* |
SDI |
= |
Standard deviation of reagent blank intensity (Io) |
C |
= |
Concentration of the calibration standard in µg/mL |
S |
= |
Solution volume in mL |
I |
= |
Total intensity of standard containing concentration C |
Io |
= |
Background intensity (reagent blank) |
|
* |
In most cases K=2 or 3 for qualitative and K=10 for quantitative
determinations. |
6.7.2. Analysis using an automatic sampler is described:
- Fill automatic sample vials with the minimum sample volume needed for one
analysis and a potential rerun.
- Load the automatic sampler with labeled standard and sample vials. A
multielement working standard should be analyzed after every 5 to 6 samples. A
control standard should be occasionally analyzed to ensure proper instrument
operation. If an element or elements contained in the control standard are not
within specification, the analyst should recalibrate before proceeding with the
analysis. A general rule is to use a value less than ±10 or 15% of the known
concentration. See Table 2 for additional details.
- Aspirate each sample or standard for approximately 1 minute prior to initiating
the exposure cycle. This ensures equilibration in the plasma and minimizes
carry-over effects.
- Dilute and reanalyze any samples containing elements (both screened and
validated) exceeding the working range. Interelement corrections may not be
accurate above the working range. Prepare the dilutions by pipetting an
appropriate aliquot from the original solution and dilute with 32% HCl/4%
HNO3.
- Based on the calibration curve initially obtained, convert the sample intensities to
concentrations. Then, using the air volume, solution volume, dilution factor, and
sample weight, calculate the concentration for each element analyzed as mg/m3
(air samples), total micrograms (wipes), or percentage of total weight (bulks)
using the equations shown below. Such calculations are usually performed by
software programs within the instrument's dedicated computer or by an external
data reduction method.
6.8. Calculations
6.8.1. The amount of analyte in each sample or blank is calculated as:
µg A |
= |
(µg/mL A) × (mL S) × (DF) |
(1) |
|
Where: |
µg A |
= |
Total µg of analyte in the sample or blank |
|
µg/mL A |
= |
Measured concentration of analyte in solution (derived from calibration curve) |
|
mL S |
= |
total volume of the solution analyzed |
|
DF |
= |
amount of dilution applied to an aliquot of the original solution
(ratio of final volume divided by the aliquot volume) |
6.8.2. The blank value, if any, is subtracted from each sample:
µgc A |
= |
µg A - µgb A |
(2) |
|
Where: |
µgc A |
= |
µg of analyte, blank corrected |
|
µg A |
= |
µg of analyte from equation (1) above |
|
µgb A |
= |
µg of analyte in blank |
6.8.3. For air samples, the concentration of analyte in the sample is expressed in
mg/m3 analyte for each element or compound analyzed.
mg A/m3 = |
(µgc A) × (GF) air volume, L |
|
(3) |
|
Where: |
GF |
= |
Gravimetric Factor |
For those elements having a PEL listed as an oxide, the gravimetric factors are:
1.2447 for ZnO,
1.4298 for Fe2O3,
1.7852 for V2O5
6.8.4. Convert bulk sample analytes to % composition using:
analyte % (w/w) = |
(µgc A) (100%) (sample weight) (1,000 µg/mg) |
|
(4) |
Where: |
µgc A |
= |
analyte amount (µg) |
|
Sample wt |
= |
aliquot (in mg) of bulk taken in Section 6.5.3. |
7. Reporting Results
7.1. Wipe sample concentrations are calculated and reported as total micrograms
(µgc A) or milligrams for each element.
7.2. Bulk sample results are calculated and reported as elemental percent by weight. Although the acid
matrix is constant for samples and standards, the differences in the overall sample matrix between
a bulk and standard can be large. Therefore, bulk results are reported as approximate values for
each element determined.
7.3. Air sample results are reported in units of mg/m3. Results for analytes
having an oxide PEL are reported as mg/m3 of the oxide.
7.4. Determinations of the screened elements or compounds are not routinely reported. Spectral
interference corrections for these analytes are not included, and validations have not been
performed. If a sample has a screened analyte over the PEL, the analyst should contact her/his
supervisor. Additional sampling or, if possible, additional analysis of the original sample should
be performed to quantitate the potential overexposure.
8. References
8.1. Occupational Safety and Health Administration Technical Center: ICP Backup Data Report
(ID-206 Solders) by D.C. Cook. Salt Lake City, UT. 1991.
8.2. National Institute for Occupational Safety and Health: NIOSH Manual of Analytical Methods,
2nd ed., Vol. 7 (DHEW/NIOSH Pub. No. 82-100). Cincinnati, OH, 1981. Method No. 351.
8.3. Occupational Safety and Health Administration Technical Center: ICP Standard Operating
Procedure by J.C. Septon. Salt Lake City, UT. 1988 (unpublished).
Table 1Calibration Standards and Wavelengths
|
NO |
|
ELE |
STD SOLN |
CONCN |
WAVELENGTH (nm) |
|
1 |
Ag |
1 |
1.00 |
328.0 |
2 |
Be |
1 |
1.00 |
313.0 |
3 |
Cd |
1 |
10.00 |
226.5 |
4 |
Pb |
1 |
10.00 |
220.3 |
|
5 |
Cu |
2 |
1.00 |
324.7 |
6 |
Sb |
2 |
10.00 |
217.5 |
7 |
Sn |
2 |
10.00 |
189.9 |
8 |
Zn |
2 |
10.00 |
213.8 |
|
9 |
Al |
3 |
10.00 |
396.1 |
10 |
As |
3 |
10.00 |
189.0 |
11 |
Ca |
3 |
10.00 |
393.3 |
12 |
Co |
3 |
10.00 |
237.9 |
13 |
Cr |
3 |
10.00 |
267.7 |
14 |
Fe |
3 |
100.00 |
271.4 |
15 |
Mg |
3 |
10.00 |
279.5 |
16 |
Mn |
3 |
10.00 |
257.6 |
17 |
Mo |
3 |
10.00 |
202.0 |
18 |
Ni |
3 |
10.00 |
231.6 |
19 |
Se |
3 |
10.00 |
196.0 |
20 |
Si |
3 |
10.00 |
212.4 |
21 |
V |
3 |
10.00 |
310.2 |
Where: ELE STD SOLN CONCN |
Element Number of Calibration standard
Concentration (µg/mL) of calibration standard |
Notes: STD SOLN 1 - Prepare in an amber-colored glass
bottle to protect the Ag from photo-decomposition.
STD SOLN 3 - This standard is used only for calibration of
the screened elements. The digestion and analysis are not optimized for these elements.
For the ISA JY-32 ICP, calibration is accomplished using a two-point calibration curve with the concentration for each
element listed above. A reagent blank was used as the low standard. Each element calibrated is contained in one of
three separate calibration standards (STD SOLN). For example, STD SOLN 1 contains Ag, Be, Cd, and Pb.
The three mixed calibration standards were selected because of chemical compatibility and potential interferences.
Other combinations of elements or concentrations can be used; however, compatibility and possible interferences have
to be considered when combining elements other than the mixtures listed above.
Table 2Control Standard
|
NO. |
ELE |
UPPER CONCN |
LOWER CONCN |
STD CONCN |
|
1 |
Pb |
5.75 |
4.25 |
5.0 |
2 |
Cd |
5.75 |
4.25 |
5.0 |
3 |
Ag |
0.575 |
0.425 |
0.5 |
4 |
Be |
0.575 |
0.425 |
0.5 |
Where: ELE UPPER CONCN LOWER CONCN STD CONCN |
Element The upper concentration limit in µg/mL
The lower concentration limit in µg/mL
The calibration concentration in µg/mL |
A control standard should be prepared and analyzed as an analytical monitor of ICP performance. Some instrument
manufacturers have instituted analytical software routines which will evaluate control standard results during the
analysis. The control standard example listed above is used for the JY-32 ICP. An alternate control
mixture can be prepared.
To illustrate the control standard concept, 5 µg/mL Pb is used. For this control standard, Pb has a limit of ± 15%
(upper and lower concentration limits of 5.75 µg/mL and 4.25 µg/mL, respectively). If a calculated
value greater than ±15% is obtained while analyzing this standard, the analysis will automatically halt.
The ICP operator should then re-calibrate the instrument.
|